The motorised collimators described
in this page are designed to be used for standard medical or veterinary radiographic investigations and are certified as attested in the individual descriptions.
All the collimators are subject to rigorous production and final
testing processes and are delivered with correct mechanical centering
of the x-ray field, with calibration of the field to ensure exact
light-field/x-ray field correspondence.
|
|
|
R302
M/A |
|
 |
| |
Motorised, square-field, x-ray collimator suitable for installation on stationary x-ray equipment. Shutter movements are motorised; shutter adjustment is possible via remote control pushbuttons
- Mounting plane: 80mm (3.14 “) from the X-ray focus.
- Without light field simulating the x-ray field.
- Radiation shielding: 150 kVp - 4mA
- Minimum inherent filtration 0,3 mm aluminium equivalent.
- Continuous field coverage from min. 00 x 00 cm a max. 48 x 48 cm ad una DFF di 100 cm
|
 |
| R302MP/A |
|
 |
| |
Motorised, square-field, x-ray collimator suitable for installation on stationary x-ray equipment. Shutter movements are motorised; shutter positioning is potentiometer controlled.
- Mounting plane: 80mm (3.14 “) from the X-ray focus.
- Without light field simulating the x-ray field.
- Radiation shielding: 150 kVp - 4mA
- Minimum inherent filtration 0,3 mm aluminium equivalent.
- Continuous field coverage from min. 00 x 00 cm a max. 48 x 48 cm ad una DFF di 100 cm
-
|
 |
| R302
ML/A |
|
 |
 |
Square field collimator for operation on stationary units
with fixed or rotating anode x-ray tube.
- Maximum radiation protection 150 kVp.
- With light field simulating the x-ray field controlled by an
electronic timer.
- Maximum field at 90 cm FFD: 43x43 cm with mounting plane at
80mm from the x-ray focus.
- Motorised control of shutters and remote positioning.
- Model with ABS plastic cover; colour: RAL9002
|
 |
| R302FML/A |
|
 |
| |
Square field collimator for operation on stationary units
with fixed or rotating anode x-ray tube.
- Maximum radiation protection 150 kVp.
- With light field simulating the x-ray field controlled by an
electronic timer.
- Maximum field at 90 cm FFD: 43x43 cm with mounting plane at
80mm from the x-ray focus.
- Motorised control of shutters and remote positioning.
- Three filters on rotating support; manual positioning.
- Model with ABS plastic cover; colour: RAL9002
|
 |
| R302MLP/A |
|
 |
| |
Motorised, square-field, x-ray collimator suitable for installation on stationary x-ray equipment. Shutter movements are motorised; shutter positioning is potentiometer controlled and adjusted remotely or by two knobs on the collimator front panel.
- Mounting plane: 80 mm (3.14") from the x-ray focus.
- External adjustment of mirror angulation.
- High luminosity provided by a quartz iodide lamp. On request luminosity may be provided by a white power Led (RO333).
- Timer limiting projection lamp exposure time to 30 seconds thus extending lamp life by preventing overheating.
- Radiation shielding: 150 kVp - 4 mA.
- Minimum inherent filtration 2 mm aluminium equivalent (1mm su richiesta/on request) .
- Variable square-field from min. min. 00 x 00 cm to max. max. 48 x 48 cm at an FFD/SID of 100 cm.
- Option of single or double laser for Potter Bucky centering.
|
 |
| R302MLPI/A |
|
 |
| |
Round
and square-field model for operation with an Image Intensifier
and Spot Film Device on remote controlled systems using rotating
anode x-ray tubes.
- Maximum radiation protection 150 kVp.
- Light field simulating the x-ray field provided by a halogen lamp or white power Led; the light field is controlled by an
electronic timer.
- Maximum field at 100 cm FFD: 43x43 cm with mounting plane at
80mm from the x-ray focus.
- Motorised control of shutters and potentiometer controlled positioning.
- Model with ABS plastic cover; colour: RAL9002
|
 |
| R302
FMLP/A |
|
 |
 |
Square field collimator for operation on stationary and
remote controlled units with fixed or rotating anode x-ray tube.
- Maximum radiation protection 150 kVp.
- With light field simulating the x-ray field controlled by an
electronic timer.
- Maximum field at 90 cm FFD: 43x43 cm with mounting plane at
80mm from the x-ray focus.
- Motorised control of shutters and potentiometer controlled positioning.
- Three filters on rotating support; manual positioning.
- Model with ABS plastic cover; colour: RAL9002
|
 |
|
R302
MFMLP/A |
|
 |
| |
Model for operation on stationary and remote controlled units with
fixed or rotating anode x-ray tube.
- Maximum radiation protection 150 kVp.
- With light field simulating the x-ray field controlled by an
electronic timer.
- Maximum field at 90 cm FFD: 43x43 cm with mounting plane at
80mm from the x-ray focus.
- Motorised control of shutters and potentiometer controlled positioning.
- Three filters on motorised rotating support; potentiometer control
of filter position.
- Model with ABS plastic cover; colour: RAL9002
|
 |
| R503
MP/A |
|
 |
| |
Round and square-field collimator for operation with an
Image Intensifier and Spot Film Device on remote controlled systems
using rotating anode x-ray tubes.
- Maximum radiation protection 150 kVp.
- Without light field simulating the x-ray field.
- Maximum field at 100 cm FFD: 43x43 cm with mounting plane at
80mm from the x-ray focus.
- Motorised control of shutters and potentiometer controlled positioning.
- Model with ABS plastic cover; colour: RAL9002
|
 |
| R503
MLP/A |
|
 |
| |
Round
and square-field model for operation with an Image Intensifier
and Spot Film Device on remote controlled systems using rotating
anode x-ray tubes.
This collimator is certified for the US market
as R302 MLPI/A
- Maximum radiation protection 150 kVp.
- With light field simulating the x-ray field controlled by an
electronic timer.
- Maximum field at 100 cm FFD: 43x43 cm with mounting plane at
80mm from the x-ray focus.
- Motorised control of shutters and potentiometer controlled positioning.
- Model with ABS plastic cover; colour: RAL9002
|
 |
| R601
AMP |
|
 |
| |
Round
and elliptic-field collimator for operation with an Image Intensifier
on mobile C-arm systems using rotating or fixed anode x-ray tubes.
- Round field and elliptic field, 9", at Focus Image Reception
Plane of 100 cm.
- Mounting plane at 80 mm from the focus.
- Maximum x-ray protection up to 125 kVp.
- Shutter movements controlled by d.c. motors.
- Joint positioning movement of the two rectangular shutters.
- Potentiometer controlled positioning of shutters on the round
field only.
|
 |
| R621
AMP |
|
 |
 |
Round
-field collimator for operation with an Image Intensifier on
mobile C-arm systems using rotating or fixed anode x-ray tubes.
- Round field, 9", at Focus Image Reception Plane of 100
cm and mounting plane at 80 mm from the focus.
- Maximum x-ray protection up to 125 kVp.
- Shutter movements controlled by d.c. motors.
- Potentiometer control of the round-field shutters only.
|
 |
| R605
MP |
|
 |
| |
Round and elliptic-field
collimator for operation with an Image Intensifier on mobile C-arm
systems using rotating or fixed anode x-ray tubes.
- Round field and elliptic fields, 12", at Focus Image Reception
Plane of 100 cm.
- Mounting plane at 80 mm from the focus.
- Maximum x-ray protection up to 125 kVp.
- Shutter movements controlled by d.c. motors.
- Joint positioning movement of the two rectangular shutters.
- Potentiometer control of the round-field shutters only.
|
 |
| R605
FMP |
|
 |
| |
Round
and elliptic-field collimator with two filtering copper shutters
1mm thick, for operation with an Image Intensifier on mobile C-arm
systems using rotating or fixed anode x-ray tubes.
- Round field and elliptic fields, 12", at Focus Image Reception
Plane of 100 cm and mounting plane at 80 mm from the focus.
- Maximum x-ray protection up to 125 kVp.
- Shutter movements controlled by d.c. motors.
- Two filtering shutters with c.c. motors.
- Joint positioning movement of the two rectangular shutters.
- Potentiometer control of the round-field shutters only.
|
 |
| R605
DLMP |
|
 |
 |
Round
and elliptic-field collimator with two pairs of rectangular
shutters which feature independent horizontal travel and rotatory
movements, for operation with an Image Intensifier on mobile C-arm
systems using rotating or fixed anode x-ray tubes.
- Round field and elliptic fields, 12", at Focus Image Reception
Plane of 100 cm
- Mounting plane at 80 mm from the focus.
- Maximum x-ray protection up to 125 kVp.
- Shutter movements controlled by d.c. motors.
- Potentiometer controlled positioning of shutters on the round
field only.
|
 |
| R625
MP |
|
 |
| |
Round -field collimator
for operation with an Image Intensifier on mobile C-arm systems
using rotating or fixed anode x-ray tubes.
- Round field, 12", at Focus Image Reception Plane of 100
cm
- Mounting plane at 80 mm from the focus.
- Maximum x-ray protection up to 125 kVp.
- Shutter movements controlled by d.c. motors.
- Potentiometer controlled positioning of shutters.
|
 |
| R651
AMP |
|
 |
| |
Round- and elliptic-field
collimator for operation with an Image Intensifier on mobile C-arm
systems using rotating or fixed anode x-ray tubes.
- Round field, 16", at Focus Image Reception Plane of 100
cm and mounting plane at 80 mm from the focus.
- Maximum x-ray protection up to 125 kVp.
- Shutter movements controlled by d.c. motors.
- Joint positioning movements of the two rectangular shutters.
- Potentiometer control of the shutters on the round field.
|
 |
| R671
AMP |
|
 |
| |
Round-field
collimator for operation with an Image Intensifier on mobile C-arm
systems using rotating or fixed anode x-ray tubes.
- Round field, 16", at Focus Image Reception Plane of 100
cm and mounting plane at 80 mm from the focus.
- Maximum x-ray protection up to 125 kVp.
- Shutter movements controlled by d.c. motors.
- Potentiometer controlled positioning of shutters.
|
 |
| R605
/170 DASM |
|
 |
 |
Round-field
collimator with two pairs of rectangular shutters
which feature independent travel and rotatory movements, for operation
with an Image Intensifier on mobile C-arm systems using rotating
or fixed anode x-ray tubes.
- Round and elliptic fields, 12", at Focus Image Reception
Plane of 100 cm and mounting plane at 80 mm from the focus.
- Round field: from <5cm (1.97") to 30 cm (12") (±1%
FFD) - for small fields (9" and 7") the secondary collimator
must be substituted. The exact field required must be specified
in the order,
- Maximum x-ray protection up to 125 kVp.
- Shutter movements controlled step-motors.
- Electronic boards in the collimator control motor positioning
and supply a return signal via CANBus.
certificazioni |
|
CE |
|
 |
| R605
/025 DASM |
|
 |
| |
Round-field collimator with
one pair of rectangular shutters which feature
independent travel and rotatory movements, for operation with an
Image Intensifier on mobile C-arm systems using rotating or fixed
anode x-ray tubes.
- Round and elliptic fields, 12", at Focus Image Reception
Plane of 100 cm and mounting plane at 80 mm from the focus.
- Round field: from <5cm (1.97") to 30 cm (12") (±1%
FFD) - for small fields (9" and 7") the secondary collimator
must be substituted. The exact field required must be specified
in the order,
- Maximum x-ray protection up to 125 kVp.
- Shutter movements controlled step-motors.
- Electronic boards in the collimator control motor positioning
and supply a return signal via CANBus.
certificazioni |
|
CE |
|
 |