motorised x-ray collimators

 
applications

The motorised collimators described in this page are designed to be used for standard medical or veterinary radiographic investigations and are certified as attested in the individual descriptions.

All the collimators are subject to rigorous production and final testing processes and are delivered with correct mechanical centering of the x-ray field, with calibration of the field to ensure exact light-field/x-ray field correspondence.

 

R302 M/A


R302M/A motorised collimator
 

Motorised, square-field, x-ray collimator suitable for installation on stationary x-ray equipment. Shutter movements are motorised; shutter adjustment is possible via remote control pushbuttons

  • Mounting plane: 80mm (3.14 “) from the X-ray focus.
  • Without light field simulating the x-ray field.
  • Radiation shielding: 150 kVp - 4mA
  • Minimum inherent filtration 0,3 mm aluminium equivalent.
  • Continuous field coverage from min. 00 x 00 cm a max. 48 x 48 cm ad una DFF di 100 cm
    certifications
    CE
Data Sheet - pdf
R302MP/A
R302MP/A motorised collimator
 

Motorised, square-field, x-ray collimator suitable for installation on stationary x-ray equipment. Shutter movements are motorised; shutter positioning is potentiometer controlled.

  • Mounting plane: 80mm (3.14 “) from the X-ray focus.
  • Without light field simulating the x-ray field.
  • Radiation shielding: 150 kVp - 4mA
  • Minimum inherent filtration 0,3 mm aluminium equivalent.
  • Continuous field coverage from min. 00 x 00 cm a max. 48 x 48 cm ad una DFF di 100 cm
  • certifications
    CE
Data Sheet - pdf
R302 ML/A
R302ML/A motorised collimator
up

Square field collimator for operation on stationary units with fixed or rotating anode x-ray tube.

  • Maximum radiation protection 150 kVp.
  • With light field simulating the x-ray field controlled by an electronic timer.
  • Maximum field at 90 cm FFD: 43x43 cm with mounting plane at 80mm from the x-ray focus.
  • Motorised control of shutters and remote positioning.
  • Model with ABS plastic cover; colour: RAL9002
    certifications
    CE
Data Sheet - pdf
R302FML/A
r302 FML/A motorised collimator
 

Square field collimator for operation on stationary units with fixed or rotating anode x-ray tube.

  • Maximum radiation protection 150 kVp.
  • With light field simulating the x-ray field controlled by an electronic timer.
  • Maximum field at 90 cm FFD: 43x43 cm with mounting plane at 80mm from the x-ray focus.
  • Motorised control of shutters and remote positioning.
  • Three filters on rotating support; manual positioning.
  • Model with ABS plastic cover; colour: RAL9002
    certifications
    CE
Data Sheet - pdf
R302MLP/A
R302MLP/A motorised collimator
 

Motorised, square-field, x-ray collimator suitable for installation on stationary x-ray equipment. Shutter movements are motorised; shutter positioning is potentiometer controlled and adjusted remotely or by two knobs on the collimator front panel.

  • Mounting plane: 80 mm (3.14") from the x-ray focus.
  • External adjustment of mirror angulation.
  • High luminosity provided by a quartz iodide lamp. On request luminosity may be provided by a white power Led (RO333).
  • Timer limiting projection lamp exposure time to 30 seconds thus extending lamp life by preventing overheating.
  • Radiation shielding: 150 kVp - 4 mA.
  • Minimum inherent filtration 2 mm aluminium equivalent (1mm su richiesta/on request) .
  • Variable square-field from min. min. 00 x 00 cm to max. max. 48 x 48 cm at an FFD/SID of 100 cm.
  • Option of single or double laser for Potter Bucky centering.
certifications
CSA FDA
Data Sheet - pdf
R302MLPI/A
R302MLPI
 

Round and square-field model for operation with an Image Intensifier and Spot Film Device on remote controlled systems using rotating anode x-ray tubes.

  • Maximum radiation protection 150 kVp.
  • Light field simulating the x-ray field provided by a halogen lamp or white power Led; the light field is controlled by an electronic timer.
  • Maximum field at 100 cm FFD: 43x43 cm with mounting plane at 80mm from the x-ray focus.
  • Motorised control of shutters and potentiometer controlled positioning.
  • Model with ABS plastic cover; colour: RAL9002
    certifications
    CSA FDA
Data Sheet - pdf
R302 FMLP/A
R302 FMLP/A motorised collimator
up

Square field collimator for operation on stationary and remote controlled units with fixed or rotating anode x-ray tube.

  • Maximum radiation protection 150 kVp.
  • With light field simulating the x-ray field controlled by an electronic timer.
  • Maximum field at 90 cm FFD: 43x43 cm with mounting plane at 80mm from the x-ray focus.
  • Motorised control of shutters and potentiometer controlled positioning.
  • Three filters on rotating support; manual positioning.
  • Model with ABS plastic cover; colour: RAL9002
    certifications
    CE
Data Sheet - pdf

R302 MFMLP/A


R302MFMLP/A motorised collimator
 

Model for operation on stationary and remote controlled units with fixed or rotating anode x-ray tube.

  • Maximum radiation protection 150 kVp.
  • With light field simulating the x-ray field controlled by an electronic timer.
  • Maximum field at 90 cm FFD: 43x43 cm with mounting plane at 80mm from the x-ray focus.
  • Motorised control of shutters and potentiometer controlled positioning.
  • Three filters on motorised rotating support; potentiometer control of filter position.
  • Model with ABS plastic cover; colour: RAL9002
    certifications
    CE
Data Sheet - pdf
R503 MP/A
R503 MP/A motorised collimator
 

Round and square-field collimator for operation with an Image Intensifier and Spot Film Device on remote controlled systems using rotating anode x-ray tubes.

  • Maximum radiation protection 150 kVp.
  • Without light field simulating the x-ray field.
  • Maximum field at 100 cm FFD: 43x43 cm with mounting plane at 80mm from the x-ray focus.
  • Motorised control of shutters and potentiometer controlled positioning.
  • Model with ABS plastic cover; colour: RAL9002
    certifications
    CE
Data Sheet - pdf
R503 MLP/A
R503 MLP/A Collimatore motorizzato
up

Round and square-field model for operation with an Image Intensifier and Spot Film Device on remote controlled systems using rotating anode x-ray tubes.
This collimator is certified for the US market as R302 MLPI/A

  • Maximum radiation protection 150 kVp.
  • With light field simulating the x-ray field controlled by an electronic timer.
  • Maximum field at 100 cm FFD: 43x43 cm with mounting plane at 80mm from the x-ray focus.
  • Motorised control of shutters and potentiometer controlled positioning.
  • Model with ABS plastic cover; colour: RAL9002
    certifications
    CE
Data Sheet - pdf
R601 AMP
R601 AMP motorised collimator
 

Round and elliptic-field collimator for operation with an Image Intensifier on mobile C-arm systems using rotating or fixed anode x-ray tubes.

  • Round field and elliptic field, 9", at Focus Image Reception Plane of 100 cm.
  • Mounting plane at 80 mm from the focus.
  • Maximum x-ray protection up to 125 kVp.
  • Shutter movements controlled by d.c. motors.
  • Joint positioning movement of the two rectangular shutters.
  • Potentiometer controlled positioning of shutters on the round field only.
    certifications
    CE
Data Sheet - pdf
R621 AMP
R621 AMP motorised collimator
su

Round -field collimator for operation with an Image Intensifier on mobile C-arm systems using rotating or fixed anode x-ray tubes.

  • Round field, 9", at Focus Image Reception Plane of 100 cm and mounting plane at 80 mm from the focus.
  • Maximum x-ray protection up to 125 kVp.
  • Shutter movements controlled by d.c. motors.
  • Potentiometer control of the round-field shutters only.
    certifications
    CE
Data Sheet - pdf
R605 MP
R605MP motorised collimator
 

Round and elliptic-field collimator for operation with an Image Intensifier on mobile C-arm systems using rotating or fixed anode x-ray tubes.

  • Round field and elliptic fields, 12", at Focus Image Reception Plane of 100 cm.
  • Mounting plane at 80 mm from the focus.
  • Maximum x-ray protection up to 125 kVp.
  • Shutter movements controlled by d.c. motors.
  • Joint positioning movement of the two rectangular shutters.
  • Potentiometer control of the round-field shutters only.
    certifications
    CE
Data Sheet - pdf
R605 FMP
R605 FMP
 

Round and elliptic-field collimator with two filtering copper shutters 1mm thick, for operation with an Image Intensifier on mobile C-arm systems using rotating or fixed anode x-ray tubes.

  • Round field and elliptic fields, 12", at Focus Image Reception Plane of 100 cm and mounting plane at 80 mm from the focus.
  • Maximum x-ray protection up to 125 kVp.
  • Shutter movements controlled by d.c. motors.
  • Two filtering shutters with c.c. motors.
  • Joint positioning movement of the two rectangular shutters.
  • Potentiometer control of the round-field shutters only.
    certifications
    CE
Data Sheet - pdf
R605 DLMP
R605 DLMP
up

Round and elliptic-field collimator with two pairs of rectangular shutters which feature independent horizontal travel and rotatory movements, for operation with an Image Intensifier on mobile C-arm systems using rotating or fixed anode x-ray tubes.

  • Round field and elliptic fields, 12", at Focus Image Reception Plane of 100 cm
  • Mounting plane at 80 mm from the focus.
  • Maximum x-ray protection up to 125 kVp.
  • Shutter movements controlled by d.c. motors.
  • Potentiometer controlled positioning of shutters on the round field only.
    certifications
    CE
Data Sheet - pdf
R625 MP
R625 MP
 

Round -field collimator for operation with an Image Intensifier on mobile C-arm systems using rotating or fixed anode x-ray tubes.

  • Round field, 12", at Focus Image Reception Plane of 100 cm
  • Mounting plane at 80 mm from the focus.
  • Maximum x-ray protection up to 125 kVp.
  • Shutter movements controlled by d.c. motors.
  • Potentiometer controlled positioning of shutters.
    certifications
    CE
Data Sheet - pdf
R651 AMP
R651 AMP
 

Round- and elliptic-field collimator for operation with an Image Intensifier on mobile C-arm systems using rotating or fixed anode x-ray tubes.

  • Round field, 16", at Focus Image Reception Plane of 100 cm and mounting plane at 80 mm from the focus.
  • Maximum x-ray protection up to 125 kVp.
  • Shutter movements controlled by d.c. motors.
  • Joint positioning movements of the two rectangular shutters.
  • Potentiometer control of the shutters on the round field.
    certifications
    CE
Data Sheet - pdf
R671 AMP
R671 AMP
 

Round-field collimator for operation with an Image Intensifier on mobile C-arm systems using rotating or fixed anode x-ray tubes.

  • Round field, 16", at Focus Image Reception Plane of 100 cm and mounting plane at 80 mm from the focus.
  • Maximum x-ray protection up to 125 kVp.
  • Shutter movements controlled by d.c. motors.
  • Potentiometer controlled positioning of shutters.
    certifications
    CE
Data Sheet - pdf
R605 /170 DASM
R605 ADLSM
su

Round-field collimator with two pairs of rectangular shutters which feature independent travel and rotatory movements, for operation with an Image Intensifier on mobile C-arm systems using rotating or fixed anode x-ray tubes.

  • Round and elliptic fields, 12", at Focus Image Reception Plane of 100 cm and mounting plane at 80 mm from the focus.
  • Round field: from <5cm (1.97") to 30 cm (12") (±1% FFD) - for small fields (9" and 7") the secondary collimator must be substituted. The exact field required must be specified in the order,
  • Maximum x-ray protection up to 125 kVp.
  • Shutter movements controlled step-motors.
  • Electronic boards in the collimator control motor positioning and supply a return signal via CANBus.
    certificazioni
    CE
Data Sheet - pdf
R605 /025 DASM
R605 ADLSM
 

Round-field collimator with one pair of rectangular shutters which feature independent travel and rotatory movements, for operation with an Image Intensifier on mobile C-arm systems using rotating or fixed anode x-ray tubes.

  • Round and elliptic fields, 12", at Focus Image Reception Plane of 100 cm and mounting plane at 80 mm from the focus.
  • Round field: from <5cm (1.97") to 30 cm (12") (±1% FFD) - for small fields (9" and 7") the secondary collimator must be substituted. The exact field required must be specified in the order,
  • Maximum x-ray protection up to 125 kVp.
  • Shutter movements controlled step-motors.
  • Electronic boards in the collimator control motor positioning and supply a return signal via CANBus.
    certificazioni
    CE
Data Sheet - pdf
 
   
 
   